Technical Specifications

Form Factor
Benchtop

High end wavelength limit
1630 nm

Wavelength Resolution
160.0fm

Instrument Type
160.0fm

Extra Specifications
Ultra high resolution Optical Spectrum Analyzer


Product Description

AP1040 Main frame equiped with OGA1040 module and OPI1040 module can achieve optical spectrum analysis with a resolution of 0.16pm. The OGA1040 module is based on an intereferometric principle, OPI1040 is giving the polarization independence. AP1040 and OGA1040 can be combined with OCA1040 module for complex spectrum analysis (chirp, phase, short pulse analysis).

  • Wavelength measurement range 1520 to 1630nm
  • Wavelength absolute accuracy +/-3pm
  • Wavelength resolution (at 3dB) 20MHz (0.16pm) and 100MHz (0.8pm)
  • Measurement level range -70dBm to +20dBm
  • Absolute level accuracy +/-0.5dB
  • Level repeatability (Typ.) +/-0.3dB
  • Close-in dynamic range >40dB @ +/-2pm; 60dB @ +/-10pm
  • Sweep time 9s for 110nm
  • Tunable laser output >-7dBm
  • Internal absolute wavelength calibrator Yes